KIRURŠKI POSTOPEK - PROVI

 


 

Na straneh, ki sledijo je korak za korakom opisan kirurški protokol za vstavitev implantata Myriad-Provi™  D2.5 L13.


 

Initial access of the osteotomy site is made using the D2.0mm pilot drill.

The pilot drill is used to establish depth and axis/direction  of drilling.

Recommended drill speeds with the pilot drill is between 800-1000 RPM.

 

 

The implant blister is peeled open to expose the Myriad-Provi™ implant that is housed in a titanium vial.  The Provi insertion adapter can be inserted into the torque ratchet or directly into the handpiece with which the implant can be directly lifted off from the packaging and delivered to the osteotomy site. 

 

The Implant can be inserted/turned into place using the torque ratchet or the handpiece.  In case the handpiece is used it must be inserted upto the top microthread using the micromotor. 

 

Final tightening is recommended to be done by the torque ratchet with an optimal insertion torque of between 35-45Ncm. 

 

After final tightening, a provisional crown can cemented on the Myriad-Provi™  implant.